Time: 2022-05-09 Author: 铧正 Click: 1325 Time
Oil Cup Cleaning Method And Common Fault Clearances
1. Oil Cup Cleaning Method
⑴ Wipe the electrode surfaces and bars again and again with clean silk cloth.
⑵ Adjust the electrode distance with standard gauge.
⑶ Use petroleum ether (other organic solvents are forbidden) to clean thrice. Each time shall follow the bellow procedures:
①Pour the petroleum ether into the oil cup till the cup is 1/4~1/3 full.
② Cover the cup mouth with a piece of glass cleaned by petroleum ether. Shake the cup evenly for one minute with certain force.
③Pour away the petroleum ether and dry the cup with a blower for 2~3 minutes.
⑷ Use the oil sample to be tested to clean the cup for 1~3 times.
①Pour the petroleum ether into the oil cup till the cup is 1/4~1/3 full.
②Cover the cup mouth with a piece of glass cleaned by petroleum ether. Shake the cup evenly for one minute with certain force.
③Pour away the left oil sample and then the test starts.
2. Agitating Blade Cleaning Method
⑴ Wipe the agitating blade again and again with clean silk cloth until fine particles are not found on their surfaces. It is forbidden to touch the surfaces with hands.
⑵ Use forceps to clamp the blade; put them into petroleum ether and wash.
⑶ Use forceps to clamp the blade and dry them with a blower.
⑷ Use forceps to clamp the blade; put them into the oil sample to be tested and wash.
3. Oil Cup Storage
Method 1 Fill the cup with good insulating oil after the test is finished and place it stable.
Method 2 Clean and dry the cup under the above procedures and then put it into a vacuum dryer.
Note: The oil cup and agitating blade shall be cleaned under the above procedures after the first test and tests with poor oil.
4、Common Fault Clearances
⑴ Power light off, screen display off
① Check the plugging of power plug.
② Check the condition of the protector tube inside the power plug.
③ Check the socket electricity.
⑵ No punch through oil cup
① Check inserting of connectors on circuit board.
② Check contacting of cap high-voltage switch.
③ Check attracting of high-voltage contacts.
④ Check break of high-voltage line.
⑶ Light display contrast
① Adjust the potentiometer on the circuit board.
⑷ Printer failure
① Check plugging of printer power line.
② Check plugging of printer data line.