在线客服系统
[sort:name]

Home >> News

Precautions and troubleshooting of the tension tester

Time: 2021-12-20 Author: 铧正 Click: 1033 Time

Intermolecular Forces formed liquid surface tension.Numbers of tension  reflect on liquid sample’s physics and chemical properties, it is one of the most important index to product quality. 

2.jpg

automatic oil interfacial tension tester

I.Matters needing attention

1. Working environment:

The temperature of the test environment should be kept constant. Do not place the instrument in an environment with high humidity and corrosive gases.

2. Platinum ring:

The plane of the ring should be parallel to the page on the north side, and the ring should have a certain roundness. The platinum ring should be clean, and it can be cleaned with detergent, and then rinsed with pure water, and then the platinum wire is heated in the oxidation flame of the alcohol lamp to orange-red.

3. Test cup:  

The test cup should be clean, and it can be cleaned and washed, then rinsed with hot water, and finally rinsed with pure water, and drained before use.

4. Obtaining pure water:

It is best to use distilled water that has been purified multiple times. Tests have shown that some commercially available purified drinking water can meet the experimental requirements.

5. Calibration of the instrument:

After the instrument is subjected to a large impact, or after moving, the weight in the setting items should be calibrated.

II.Fault check and treatment


Phenomenon

Reason

Disposal methods

1

The weight cannot be calibrated

The polarity of the magnetic cylinder is wrong

Stick to the correct polarity    Replace the magnetic cylinder

2

LCD screen does not display

LCD cable is not plugged in

Plug in the LCD screen cable

3

No response at boot

The switch is not in good contact

Blown fuse

Replace the switch Replace the fuse

4

Test data is too small

Parameter setting is incorrect           Platinum ring is not clean

Set the parameters correctly          Clean the platinum ring

5

The data is abnormal

Weight calibration is incorrect

Recalibrate the weight


+8618330222305 +8618330222305 sales08@bdhuazheng.com 0086-18330222305 Julia Jin-huazheng